MLS-3020U
External Dimensions (W x D x H) | 440 x 550 x 1050 mm |
---|---|
Chamber Dimensions (diameter x depth) | 300 x 670 mm |
Effective Chamber Height Incl. Recess in Lid | – |
Effective Capacity | 47 litres |
Net Weight | 67 kg |
Material | – |
Chamber Material | Stainless steel (SUS 304) |
Interior Finish | – |
Max. Pressure | – |
Sterilisation Temperature | 105 ~ 126 °C |
Temperature Gauge Range | Digital display 80 ~ 141 °C |
Culture Media Melting Temperature | – |
Keep Warm Temperature | – |
Safety Valve Release Pressure | 177 kPa (25 psig) |
Pressure Gauge Range | 0 – 0.3 Mpa / 0 -45 psi |
Timer Setting Range | 1 – 180 min |
Sterilisation Timer | – |
Melting Timer | – |
Keep Warm Timer | – |
Exhaust Tank | 3 ltr polyethylene tank |
Exhaust Control | – |
Pressure Safety Valve | Y |
Over-Temperature Limiter | – |
Over-Pressure Limiter | – |
Anti-Scorch Limiter | Y |
Door Interlock | – |
Current Fuse | Y |
Pressure Vessel | – |
MLS-3751L
External Dimensions (W x D x H) | 478 x 632 x 748 mm |
---|---|
Chamber Dimensions (diameter x depth) | 370 x 415 mm |
Effective Chamber Height Incl. Recess in Lid | 463 mm |
Effective Capacity | 50 litres |
Net Weight | 61 kg |
Material | – |
Chamber Material | Stainless steel (SUS 304) |
Interior Finish | – |
Max. Pressure | 0.240 MPa |
Sterilisation Temperature | 115 ~ 135 °C |
Temperature Gauge Range | – |
Culture Media Melting Temperature | 60 ~ 114 °C |
Keep Warm Temperature | 45 ~ 60 °C |
Safety Valve Release Pressure | – |
Pressure Gauge Range | – |
Timer Setting Range | – |
Sterilisation Timer | 1 to 300 min |
Melting Timer | 1 to 300 min |
Keep Warm Timer | 72 hours fixed |
Exhaust Tank | 2 ltr polyethylene tank |
Exhaust Control | Exhaust valve open temperature setting |
Pressure Safety Valve | Y |
Over-Temperature Limiter | Y |
Over-Pressure Limiter | Y |
Anti-Scorch Limiter | Y |
Door Interlock | Y |
Current Fuse | Y |
Pressure Vessel | Small-scale pressure vessel |
MLS-3781L
External Dimensions (W x D x H) | 478 x 632 x 965 mm |
---|---|
Chamber Dimensions (diameter x depth) | 370 x 630 mm |
Effective Chamber Height Incl. Recess in Lid | 688 mm |
Effective Capacity | 75 litres |
Net Weight | 71 kg |
Material | Stainless steel (SUS 304) |
Chamber Material | Stainless steel plate |
Interior Finish | Stainless steel |
Max. Pressure | 0.240 MPa |
Sterilisation Temperature | 115 ~ 135 °C |
Temperature Gauge Range | – |
Culture Media Melting Temperature | 60 ~ 114 °C |
Keep Warm Temperature | 45 ~ 60 °C |
Safety Valve Release Pressure | – |
Pressure Gauge Range | – |
Timer Setting Range | – |
Sterilisation Timer | 1 to 300 min |
Melting Timer | 1 to 300 min |
Keep Warm Timer | 72 hours fixed |
Exhaust Tank | 2 ltr polyethylene tank |
Exhaust Control | Exhaust valve open temperature setting |
Pressure Safety Valve | Y |
Over-Temperature Limiter | Y |
Over-Pressure Limiter | Y |
Anti-Scorch Limiter | Y |
Door Interlock | Y |
Current Fuse | Y |
Pressure Vessel | Small-scale pressure vessel |